Stentectomy by an Entrapped Rotational Atherectomy Burr

Video 1. Coronary angiography demonstrated multiple severe and heavily calcified stenoses in the mid- and distal left anterior descending artery in previously implanted stents.

Video 2. Upon withdrawal, the burr became entrapped in the mid-left anterior descending artery within a stented segment.

Video 3. Advancement of a guide catheter extension over the burr into the mid-left anterior descending artery.  

Video 4. After multiple attempts and pulling the guidewire, the entrapped burr was removed.

Video 5. Angioplasty was performed with intravascular lithotripsy and non-compliant balloons.


Video Supplement to "Stentectomy by an Entrapped Rotational Atherectomy Burr" (Clinical Image).