Skip to main content

Advertisement

ADVERTISEMENT

Clinical Images

Dual-Lumen Catheter and Floating-Wire Technique to Access Protruding Aorto-Ostial Stent

Bernard Wong, MBChB and Eugene B. Wu, MD, DM

May 2023
1557-2501
J INVASIVE CARDIOL 2023;35(5):E275-E276. doi: 10.25270/jic/22.00289

J INVASIVE CARDIOL 2023;35(5):E275-E276

Key words: right coronary artery, stent protrusion

A 56-year-old man with exertional angina underwent coronary angiography, which showed severe restenosis within a previously placed stent in the ostial right coronary artery (RCA). Selective engagement of the RCA with a JR4 diagnostic catheter was difficult due to excessive protrusion of the previous stent into the aorta (Figure 1A). A 6-Fr, JR4 guide catheter was used for intervention.

A workhorse guidewire was able to enter the conus branch through a protruding side-strut. Wiring the RCA proved difficult and could only be done through a protruding side-strut (Figure 1B). A Sasuke dual-lumen catheter (DLC; Asahi Intecc) was loaded onto the RCA guidewire, but the central stent lumen could not be wired successfully (Figure 1C). The DLC was then loaded onto a floating wire, and the central stent lumen was successfully accessed with a second guidewire (Figure 1D; Video Series).

Wong Floating-Wire Technique Figure 1
Figure 1. (A) Baseline angiogram showing significant ostial right coronary artery restenosis with difficult selective engagement due to protrusion of previous stent. (B) The conus branch and RCA could only be wired through a protruding side strut. (C) A Sasuke dual lumen catheter was loaded onto the guidewire through the side strut with floating guidewire support. Another guidewire in the OTW lumen could only also wire the protruding side strut. (D) The Sasuke dual lumen catheter was moved onto the floating guidewire and the central lumen of the protruding stent was successfully wired.

Intravascular ultrasound (IVUS) confirmed wiring of the central lumen, with 4 mm of protrusion into the aorta. The ostial RCA had a minimal lumen area of 3.2 mm2, likely due to mechanical compression at the aorto-ostial junction (Figure 2A; Video Series). A 3.5- x 14-mm drug-eluting stent was deployed and postdilated with a 4.0-mm non-compliant balloon at high pressure. IVUS showed minimal stent area of 6.8 mm2, with stent expansion of 93% (Figures 2B and 2C; Video Series).

Wong Floating-Wire Technique Figure 2
Figure 2. (A) Intravascular ultrasound showing restenosis in the ostial right coronary artery with minimal lumen area of 3.2 mm2. (B) Minimal stent area of 6.8 mm2. (C) Distal stent reference area of 7.3 mm2. Stent expansion of 93%.

Percutaneous coronary intervention in the setting of previous aorto-ostial stenting can be difficult, especially if there is excessive stent protrusion. Various techniques have been described, including double-wire technique, double-guide snare technique, side-strut sequential ballooning technique, and guide-extension facilitated side-strut stenting.1-5 These techniques can sometimes be complicated, and intervention through a side-strut may lead to excessive stent deformation or avulsion of the protruding segment. Our novel technique uses a DLC and floating wire to back the JR4 guide away from the protruding stent while maintaining stability for another guidewire to enter the central lumen.

Affiliations and Disclosures

From Prince of Wales Hospital, Chinese University of Hong Kong, Hong Kong SAR, China.

Disclosure: The authors have completed and returned the ICMJE Form for Disclosure of Potential Conflicts of Interest. Dr Wu reports research funding from OrbusNeich, Asahi Intecc, Abiomed; consulting honoraria from Boston Scientific and Abbott Vascular; member of the board of directors for the APCTO club. Dr Wong reports no conflicts of interest regarding the content herein.

The authors report that patient consent was provided for publication of the images used herein.

Manuscript accepted September 15, 2022.

Address for correspondence: Bernard Wong, MBChB, Division of Cardiology, Department of Medicine and Therapeutics, Prince of Wales Hospital, Chinese University of Hong Kong, Hong Kong SAR, China. Email: bernardwong@hotmail.co.nz

References

1. Chetcuti SJ, Moscucci M. Double-wire technique for access into a protruding aorto-ostial stent for treatment of in-stent restenosis. Catheter Cardiovasc Interv. 2004;62(2):214-217. doi:10.1002/ccd.20062

2. Uehara Y, Shimizu M, Yoshimura M. A novel technique for catheter engagement of protruding aorto-ostial stent. Catheter Cardiovasc Interv. 2014;83(7):1093-1096. doi:10.1002/ccd.25274

3. Burstein JM, Hong T, Cheema AN. Side-strut stenting technique for the treatment of aorto-ostial in-stent restenosis and deformed stent struts. J Invasive Cardiol. 2006;18(8):E234-E237.

4. Esenboga K, Sahin E, Ozyuncu N, Yamanturk Y, Turhan S. Challenging intervention to restenosis of right coronary ostial stent excessively overhanging to the aorta: a case report and brief review of literature. Cureus. 2022;14(5):e25037. doi:10.7759/cureus.25037

5. Kassimis G, Raina T. GuideLiner extension catheter-facilitated side strut stenting technique for the treatment of right coronary artery ostial in-stent restenosis. Cardiovasc Revasc Med. 2018;19(1 Pt B):133-136. doi:10.1016/j.carrev.2017.09.010


Advertisement

Advertisement

Advertisement