Original Contribution

Automated Carbon Dioxide Digital Angiography for Lower-Limb Arterial Disease Evaluation: Safety Assessment and Comparison With Standard Iodinated Contrast Media Angiography

Filippo Scalise, MD1;  Eugenio Novelli, PhD2;  Carla Auguadro, MD1;  Valentina Casali, MD1;
Mariella Manfredi, MD1;  Romano Zannoli, PhD, EngD3

January 2015

Abstract: Introduction. Carbon dioxide (CO2) has been validated as a contrast agent in a large series of studies. A particular advantage of CO2 over iodinated contrast medium (ICM) is the absence of nephrotoxicity and allergic reactions. One of the limitations of CO2 angiography is the difficulty of CO2 manual injection due to its compressibility. The manual gas injection does not permit optimal control of the gas output. Development of an automated CO2 injector has overcome these problems. Aim. This study compares the feasibility, safety, and diagnostic accuracy of automated CO2 digital subtraction angiography (DSA) in comparison with ICM-DSA in the evaluation of critical limb ischemic (CLI) patients. Methods. We performed DSA with both CO2 and ICM on 40 consecutive CLI patients and directly compared the two techniques. Sixteen females and 24 males participated in the study (mean age, 71.7 years). We assessed the diagnostic accuracy of CO2 in identifying arterial stenosis in the lower limb, with ICM-DSA used as the gold standard. Results. The overall diagnostic accuracy of CO2-DSA was 96.9% (sensitivity, 99.0%; specificity, 96.1%; positive predictive value, 91.1%; negative predictive value, 99.6%). Tolerable minor symptoms occurred in 3 patients. No allergic reactions or significant decline in renal function were observed in patients receiving the CO2 injection. Conclusion. Carbon dioxide DSA is a valuable and safe alternative to traditional ICM-DSA for evaluating CLI patients. This modality should be considered as the standard choice for CLI patients undergoing angiographic evaluation who are known to have renal insufficiency or contrast allergy.

J INVASIVE CARDIOL 2015;27(1):20-26
Key words: carbon dioxide, angiography, peripheral vascular disease, contrast media

_____________________________________________________________

 

Carbon dioxide (CO2) has been used for diagnostic purposes in humans since the early 1900s.1,2 It is currently used as an alternative to iodinated contrast media (ICM) in several diagnostic and interventional procedures.3-8 Studies have shown potential benefits in the use of CO2 over ICM for digital subtraction angiography (DSA) vascular imaging.3-5 Carbon dioxide is an inexpensive gas that is widely available, providing cost savings compared with traditional ICM. It is highly soluble in blood (20 times more soluble than oxygen) and is rapidly eliminated by the lungs in the first pass, allowing the injection of almost unlimited quantities of gas provided that adequate time for elimination is allowed between injections.9 Carbon dioxide has no known inherent nephrotoxicity, making it desirable for evaluating patients with evidence of renal dysfunction.3,5,9,10 Additionally, there is no potential for allergic contrast reactions. Because of the low viscosity of CO2 (400 times less than ICM), smaller angiographic catheters may be used, and the filling of severely diseased stenotic vessels may be enhanced.11 During the injection, the gas displaces the column of blood and acts as a negative contrast agent. The small change in density between a blood vessel containing blood and a blood vessel containing gas can be demonstrated using DSA. The C02 x-ray absorption is roughly one-tenth of the absorption obtained with diluted iodine.4,11,12 For these reasons, both the injection process and the radiologic set-up must be optimized to yield good-quality angiographic images.13-15 Various authors have described their experience with catheter-based angiography using CO2 as the radiographic contrast agent for the evaluation of arterial and venous diseases. These reported uses have included diagnostic evaluations of the abdominal8,13-15 and extremity vessels,6,16-18 renal transplants,19 tumors,20 and hemodialysis access sites.7 Therapeutic interventions such as balloon angioplasty,21,22 stent placement,18,21,22 caval interruption procedures,23,24 and transjugular intrahepatic portosystemic shunts25 using CO2 as the contrast agent have also been reported. Despite these encouraging results, the use of CO2 as a contrast agent in the evaluation and treatment of peripheral arterial disease (PAD) has not been as prevalent as expected. There are various explanations for this occurrence. The use of a gas rather than a liquid contrast agent changes some steps in the process of angiography that may be unfamiliar to vascular interventionalists. The injection of a gaseous element, which is mostly performed by hand, requires a long training period, and the vascular image results are not always ideal due to the mechanical behavior and the small radiological absorption coefficient of the gaseous contrast medium.4-11,26 Expert operators insist that a long period of empirical training is required to obtain good-quality images. This training is necessary to acquire the skill to control the hand injection and to become accustomed to the natural properties of CO2 to obtain optimal gas filling of the vessels, particularly in the distribution of the vessels below the knee.16,17,26 The main limitation of CO2 use in angiography is the unreliability of the CO2 injection; its compressibility and possible contamination with air can lead to a potentially fatal complication.27 The development of an automated CO2 injector (Angiodroid, Angiodroid SRL) has addressed these problems. This automatic gas injector with low-pressure automatic line washout may limit the risk of liquid jet injury and optimize vessel imaging through an optimal regulation of the amount and flow of the injected gas.26


The aim of this study was to compare the feasibility, safety, and diagnostic accuracy of automated CO2-DSA with standard ICM-DSA in the evaluation of PAD.

Methods

Scalise Angiography Table 1
Table 1. Population characteristics.

Patients. During an 8-month period, from September 2012 to April 2013, a total of 40 CLI patients underwent lower-limb angiography with both ICM and CO2 contrast agents to directly compare the two techniques. Sixteen female and 24 male patients participated; their mean age was 71.7 years (range, 50-82 years). The baseline clinical characteristics of the treated patients are shown in Tables 1 and 2. Medical illnesses included the following: diabetes in 24 patients; chronic renal insufficiency in 31 patients; hypertension in 23 patients; stable coronary artery disease in 9 patients; mild chronic obstructive pulmonary disease (COPD, stage I GOLD classification) in 5 patients; and cerebrovascular disease in 4 patients. Exclusion criteria were: COPD stage II, III, or IV GOLD classification; severe ischemic limb ulcers or frank gangrene (category 6 Rutherford classification); atrial and ventricular septal defects; or pulmonary arteriovenous malformation. Study approval was obtained from the local ethics committee, and informed consent was obtained from each patient at the time of the investigation. The patients who consented to participate in the study had angiography performed with both CO2 and ICM to directly compare the two techniques. CO2-DSA was performed before ICM-DSA in the same procedure.

 

Scalise Angiography Table 2
Table 2. Peripheral arterial disease prevalence.

CO2 and ICM digital angiography. All patients received the same preparation they would receive for iodinated angiograms. Angiography was performed via femoral arterial puncture. CO2 was delivered using the Angiodroid CO2 injector via a straight selective 4 Fr catheter positioned at three different levels: common iliac artery, common femoral artery, and popliteal artery. The volume of gas injection was chosen considering the injection site: between 40 mL (popliteal artery) and 60 mL (iliac and femoral arteries) of CO2 were used per injection at 300-400 mm Hg of pressure. In each patient, we performed an average of 6 CO2 injections. The total average volume of CO2 used was 240 ± 80 mL. For lower-extremity procedures, elevation of the extremity often enhanced the quality of the images by taking advantage of the lower density of CO2 compared to blood. DSA images were obtained in an anteroposterior (AP) projection and in both oblique projections. At least 2 minutes were allowed between injections to ensure complete excretion of CO2. All studies were performed using the GE Innova DSA system (GE Medical Systems) with a postprocessing software for CO2 images. A high-mA (90 mA) and low-kV (70 kV) technique was used. Because of the quick passage of the CO2 through the blood stream, increased frame rates were generally required to acquire pictures (7 frames/s is a typical setting for most systems). Iodinated contrast medium angiography was performed using iodixanol 320 mg/mL (Visipaque; GE Healthcare) via 4 Fr Pigtail catheter positioned at the abdominal aortic bifurcation. Between 15-20 mL of iodixanol were used per injection and the total volume average of contrast medium used was 150 ± 50 mL. Patients were questioned at the time of the study regarding any symptoms experienced during or following the CO2 and ICM injections. Following the investigation, the patients were observed for 48 hours and any clinical incidents were recorded.


Nine-territory lower-limb angiography model. For the comparison of the two investigative methods, we used a 9-territory angiographic model. Six territories were included in the first district above the knee (ATK: common iliac artery, external iliac artery, common femoral artery, profunda femoral artery, superficial femoral artery, and popliteal artery). Three territories were included in the second district below the knee (BTK: anterior tibial artery, posterior tibial artery, and peroneal artery). Evaluation and comparison of ICM-DSA and CO2-DSA were conducted independently by two operators who did not perform the arteriography procedure. A stenosis was considered significant if it was greater than 50% by visual estimate.


Statistical analysis. Descriptive statistics (counts and percentages, means ± standard deviations) are presented for patients’ demographic and clinical characteristics. The change in serum creatinine level pre-DSA and post-DSA was evaluated with a paired 2-tailed Student’s t-test. The performance of CO2-DSA for diagnosis of significant stenosis compared with the gold-standard ICM-DSA method was determined with regard to sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), accuracy, and corresponding 95% confidence intervals (CIs). Because the values of the sample proportions were very close to 1, we did not use the normal approximation to compute CIs. The Clopper-Pearson interval was used, and the exact method was based directly on the binomial distribution. The main analysis was performed at the level of ATK, BTK, and overall districts. Secondary analyses were performed at the level of the 9 districts taken individually. The data were collected and reviewed in Microsoft Excel, and statistical analysis was performed with SPSS 13.0 (SPSS, Inc).

Results

Scalise Angiography Table 3
Table 3. Carbon dioxide angiography diagnostic accuracy in the above-the-knee, below-the-knee, and overall districts.

The two investigative methods were compared by analysis of the nine territories in the entire group of 40 patients with a total number of 360 evaluated segments. Table 3 shows an overall diagnostic accuracy of 96.9% for CO2-DSA, using ICM-DSA as the gold standard (sensitivity, 99.0%; specificity, 96.1%; PPV, 91.1%; NPV, 99.6%). The diagnostic accuracy was 97.1% in the ATK district (sensitivity, 100%; specificity, 96.2%; PPV, 88.7%; NPV, 100%) and 96.7% in the BTK district (sensitivity, 97.9%; specificity, 95.8%; PPV, 94.0%; NPV, 98.6%). The diagnostic accuracy values of CO2-DSA in the 9 districts taken individually are shown in Table 4.

 

Scalise Angiography Table 4
Table 4. Carbon dioxide digital subtraction angiography diagnostic accuracy in the 9-district model.

No significant decline in renal function was observed in patients with a normal basal creatinine value (average, 0.10 mg/dL), but the average increase in serum creatinine level was 0.24 mg/dL (P<.001) in patients who were considered most at risk (serum creatinine level ≥2 mg/dL). Tolerable minor symptoms, including foot pain, occurred in 3 patients, and 1 patient experienced nausea. No cardiovascular events were noted while monitoring the patients during and after the CO2 injections. No life-threatening complications occurred during the use of CO2-DSA. No allergic reactions were noted following CO2 injection, although 3 patients experienced reversible cutaneous erythema after ICM injection.


Discussion

Carbon dioxide has several attractive properties as an intravascular contrast agent. It is non-allergenic, eliminating the possibility of fatal hypersensitivity reactions.27-29 Allergic reactions requiring some form of medical intervention may complicate up to 5% of all conventional angiography investigations.28,29 Both ionic and non-ionic ICM can cause fatal allergic reactions.30 Pretreatment with high-dose corticosteroids can prevent contrast-related anaphylactic reactions, but CO2-DSA eliminates the need for such pretreatment. In this study, 3 patients (7.5%) experienced diffuse cutaneous erythema after ICM injection, which resolved with the administration of intravenous corticosteroids.


In addition, there is no evidence in either clinical experience or animal studies to suggest that CO2 is nephrotoxic.31 Deterioration in renal function following ICM arteriography has been reported to occur in up to 11% of all patients; most of these patients are diabetic, with laboratory evidence of renal insufficiency.32,33 Iodinated contrast medium arteriography led to the need for permanent dialysis in 8% of patients with a serum creatinine >1.8 mg/dL or blood urea nitrogen >30 mg/dL.34 Because chronic renal insufficiency with or without diabetes is present in many patients with symptomatic PAD, the risk of inducing renal failure in this patient population has precluded the use of arteriography.

Although the advantage of using CO2-DSA in patients with renal dysfunction is clearly demonstrated in the literature, and although patient hydration, bicarbonate infusion, and N-acetylcysteine have been used in an attempt to reduce the nephrotoxic effects of iodinated contrast, contrast nephrotoxicity remains an issue.35,36 In this study, no significant decline in renal function was found in patients with a normal basal creatinine value (average, 0.10 mg/dL), but the average increase in serum creatinine level was 0.24 mg/dL (P<.001) in patients considered most at risk (serum creatinine level ≥2 mg/dL). Because both contrast agents were administered in the same session, the increase in serum creatinine levels was attributed to the use of ICM, since there is no evidence in clinical experience or animal studies of nephrotoxicity associated with CO2 use.


The properties of CO2 allow it to be used effectively in patients with renal failure and patients who are allergic to ICM. These properties obviate the need for preangiography hydration in patients in whom cardiac and renal (or other metabolic) dysfunctions coexist. CO2 can be used in sequential studies on consecutive days, as is frequently required for completion of endovascular procedures.


Many interventionalists expressed concern about the adequacy of imaging using CO2 in the lower extremities, particularly in smaller BTK vessels. Several studies have reported the use of CO2 as an alternative to ICM agent with a good diagnostic quality in lower-limb arteries located ATK.37-39 In a study by Rolland et al, the imaging quality of CO2-DSA was comparable to ICM-DSA at the pelvis in 93% and at the thigh in 74% of 120 arteries studied.39 The same quality was achieved distally in only half of the cases. Oliva et al found no significant differences in the mean stenosis values obtained with CO2 or ICM in any segment for any of the observers;37 however, imaging of BTK vessels using CO2 has not been reported to show such favorable results.40-42 In our series, CO2 angiography showed a diagnostic accuracy of 97.1% in the ATK district and 96.7% in the BTK district (Table 1). There are several reasons for these results: the prevalence of patients with a complete occlusion of the proximal vessels, the site of the CO2 injection, the use of the CO2 delivery system, and the use of CO2-dedicated postprocessing software.


In the referenced studies, inadequate opacification of the BTK vessels by CO2 could be related to the high prevalence of patients with complete occlusion of the proximal vessels, which causes slow distal flow. In patients in whom aortic injection of CO2 was used for bilateral lower-limb evaluation, there was a fragmentation of the CO2 gas column, which degrades the image quality, especially in the BTK arteries.17 In our study, the CO2 was administered selectively in the lower limbs and in the presence of a complete artery occlusion by placing the catheter proximal and as close as possible to the occlusion. Several authors16,39 have suggested selective arterial injection in cases of suboptimal opacification due to fragmentation. Hawkins and Caridi suggested that the catheter should be placed as close as possible to the lesion to improve the filling of the vessel with the gas.43


Another issue to consider is that CO2 was delivered intravascularly by a hand injection with a 50 mL syringe in most studies. Because CO2 gas in compressible syringes is loaded under pressure from CO2 cylinders, these syringes contain indeterminate amounts of CO2. In contrast to liquid injections, in which the rate of delivery remains constant, the rate of CO2 delivery increases exponentially toward the end of injection.9 In our opinion, this produced inconsistent CO2 delivery, and consequently, poor vessel filling.


The high value of overall diagnostic accuracy of CO2-DSA evidenced by our study (96.9%) was determined by using automated injection of the gas. This automated injection allowed us to take full advantage of the physical characteristics of the gas, drastically reducing the risks associated with manual injection. The automated CO2 angiography was safer due to the internal circuit of the injector, which maintained positive pressure to prevent the introduction of air from outside the system. The CO2 procedure was fully automatic, repeatable, and independent of the operator. Once the volume and pressure parameters were set up, the injector automatically handled the injection of the gas toward the infusion line.

 

Scalise Angiography Figure 1
FIGURE 1. Below-the-knee imaging comparing (A) carbon dioxide (CO2) angiography and (B) iodinated contrast media (ICM).

As a final point, the use of CO2 as an intraarterial contrast agent also required the use of digital subtraction technology. This technology detects a very low concentration of contrast agent by subtracting the presence of soft tissue before contrast injection and enhancing the postcontrast images through manipulation and amplification of a digitized radiographic image. Current angiography procedures are now set up with CO2 postprocessing “stacking” software. CO2 produces contrast by causing complete displacement of the blood column in the vessel. In each x-ray acquisition, the CO2 will sometimes fail to opacify the entire field during the same run. For postprocessing visualization of the entire vasculature, “stacking” techniques are used to stack the individual images on top of each other to form a single composite image.9 The ability to stack these images is essential for obtaining a final picture that delineates the true anatomy. By employing some of the strategies previously mentioned, excellent imaging of the lower extremities can be obtained (Figures 1, 2, and 3).

 

Scalise Angiography Figure 2
FIGURE 2. Femoropopliteal bypass imaging comparing (A) iodinat- ed contrast media (ICM) and (B) carbon dioxide (CO2) angiography.

Finally, consideration must be given to the ability of the automated CO2 procedure to be painless for the patient. The definition and control of injection pressure allows the operator to maintain an acceptable pressure of the gas injection into the artery. Excessive pressure normally causes pain for the patient in the manual procedures, because it is basically impossible to control low pressure values. In our study, only 3 patients experienced transient foot pain after CO2 injection. CO2 as a contrast agent is very inexpensive; however, the cost of the injector is several thousand euros, and the cost of the connectors that transmit CO2 from the injector to the catheter must also be considered.

 

Scalise Angiography Figure 3
FIGURE 3. Superficial femoral artery imaging comparing (A) carbon dioxide (CO2) angiography and (B) iodinated contrast media (ICM).

As technology advances, this imaging will likely continue to improve, making CO2 arteriography a valuable tool in the armamentarium of the vascular interventionalist when evaluating the lower extremities. This approach represents a potential imaging alternative, or supplement, to standard contrast agents for patients who are candidates for an endovascular procedure but are at high risk for contrast-related complications.

Conclusion

CO2-DSA using the Angiodroid automated delivery system is a safe alternative technique for the evaluation of patients with PAD of the lower limbs. Adequate opacification of the ATK and BTK arteries can be obtained with proper injection technique. We suggest that CO2 should be used as the initial contrast agent for the evaluation of PAD in patients with renal failure and iodine contrast allergy. For infrapopliteal segment opacification, we recommend selective injection as close to the target artery as possible. To maximally optimize imaging, the proceduralist must take advantage of the special properties of CO2. This requires some changes in angiographic techniques from contrast preparation to image postprocessing.

References

  1. Rautenberg E. Rontgenphotographieder Leber, der Milz, und des Zwerchfells. Deutsche Med Wschr. 1914;40:1205-1208.
  2. Harvard BM, White RR, Walsh JF. Experimental studies in acute retroperitoneal carbon dioxide insufflation. J Urol. 1959;81(3):481-485.
  3. Hawkins IF, Cho KJ, Caridi JG. Carbon dioxide angiography to reduce the risk of contrast-induced nephropathy. Radiol Clin North Am. 2009;47(5):813-825.
  4. Hawkins IF, Wilcox CS, Kerns SR, Sabatelli FW. CO2 digital angiography: a safer contrast agent for renal vascular imaging? Am J Kidney Dis. 1994;24(4):685-694.
  5. Nadolski GJ, Stavropoulos SW. Contrast alternatives for iodinated contrast allergy and renal dysfunction: options and limitations. J Vasc Surg. 2013;57(2):593-598.
  6. Kawasaki D, Fujii K, Fukunaga M, Masutani M, Nakata A, Masuyama T. Safety and efficacy of endovascular therapy with a simple homemade carbon dioxide delivery system in patients with ileofemoral artery diseases. Circ J. 2012;76(7):1722-1728.
  7. Kariya S, Tanigawa N, Kojima H, et al. Efficacy of carbon dioxide for diagnosis and intervention in patients with failing hemodialysis access. Acta Radiol. 2010;51(9):994-1001.
  8. Lee AD, Hall RG. An evaluation of the use of carbon dioxide angiography in endovascular aortic aneurysm repair. Vasc Endovascular Surg. 2010;44(5):341-344.
  9. Back MR, Caridi JG, Hawkins IF, Seeger JM. Angiography with carbon dioxide. Surg Clin N Am. 1998;78(4):575–591.
  10. Hawkins IF, Cho KJ, Caridi JG. Carbon dioxide in angiography to reduce the risk of contrast-induced nephropathy. Radiol Clin North Am. 2009;47(5):813-825.
  11. Huber PR, Leimbach ME, Lewis WL, Marshall JJ. CO2 angiography. Catheter Cardiovasc Interv. 2002;55(3):398-403.
  12. Shaw DR, Kessel DO. The current status of the use of carbon dioxide in diagnostic and interventional angiographic procedures. Cardiovasc Intervent Radiol. 2006;29(3):323-331.
  13. Chao A, Major K, Kumar SR, et al. Carbon dioxide digital subtraction angiography-assisted endovascular aortic aneurysm repair in the azotemic patient. J Vasc Surg. 2007;45(3):451-458.
  14. Huang SG, Woo K, Moos JM, et al. A prospective study of carbon dioxide digital subtraction versus standard contrast arteriography in the detection of endoleaks in endovascular abdominal aortic aneurysm repairs. Ann Vasc Surg. 2013;27(1):38-44.
  15. Criado E, Upchurch GR Jr, Young K, et al. Endovascular aortic aneurysm repair with carbon dioxide-guided angiography in patients with renal insufficiency. J Vasc Surg. 2012;55(6):1570-1575.
  16. Bees NR, Beese RC, Belli AM, Buckenham TM. Carbon dioxide angiography of the lower extremities: initial experience with an automated carbon dioxide injector. Clin Radiol. 1999;54(12):833-838.
  17. Madhusudhan KS, Sharma S, Srivastava DN, et al. Comparison of intra-arterial digital subtraction angiography using carbon dioxide by “home made” delivery system and conventional iodinated contrast media in evaluation of peripheral arterial occlusive disease of the lower limbs. J Med Imaging Radiat Oncol. 2009;53(1):40-49.
  18. Higashimori A, Yokoi Y. Stent implantation for chronic total occlusion in the iliac artery using intravascular ultrasound-guided carbon dioxide angiography without iodinated contrast medium. Cardiovasc Interv Ther. 2013;28(4):415-418.
  19. Moresco KP, Patel NH, Namyslowski Y, et al. Carbon dioxide angiography of the transplanted kidney: technical considerations and imaging findings. AJR Am J Roentgenol. 1998;171(5):1271-1276.
  20. Wong AA, Charalel RA, Louie JD, Sze DY. Carbon dioxide contrast enhancement for C-arm CT utility for treatment planning during hepatic embolization procedures. J Vasc Interv Radiol. 2013;24(7):975-980.
  21. Pérez L, Lecannelier E, Fernández A, Parra J, Aburto G, Robles I. Renal angioplasty with stent using CO2 as contrast medium: report of one case. Rev Med Chil. 2007;135(3):365-369. Epub 2007 Apr 26.
  22. Kessel DO, Robertson I, Patel J 5th, et al. Carbon-dioxide-guided vascular interventions: technique and pitfalls. Cardiovasc Intervent Radiol. 2002;25(6):476-483. Epub 2002 Jul 23.
  23. Dewald CL, Jensen CC, Park YH, et al. Vena cavography with CO(2) versus with iodinated contrast material for inferior vena cava filter placement: a prospective evaluation. Radiology. 2000;216(3):752-757.
  24. Boyd-Kranis R, Sullivan KL, Eschelman DJ, Bonn J, Gardiner GA. Accuracy and safety of carbon dioxide inferior vena cavography. J Vasc Interv Radiol. 1999;10(9):1183-1189.
  25. Krajina A, Lojik M, Chovanec V, Raupach J, Hulek P. Wedged hepatic venography for targeting the portal vein during TIPS: comparison of carbon dioxide and iodinated contrast agents. Cardiovasc Intervent Radiol. 2002;25(3):171-175. Epub 2002 Feb 25.
  26. Corazza I, Rossi PL, Feliciani G, Pisani L, Zannoli S, Zannoli R. Mechanical aspects of CO2 angiography. Phys Med. 2013;29(1):33-38. Epub 2011 Dec 3.
  27. Caridi JG, Hawkins IF. CO2 digital subtraction angiography: potential complications and their prevention. J Vasc Intervent Radiol. 1997;8(3):383-391.
  28. Palmiere C, Comment L, Mangin P. Allergic reactions following contrast material administration: nomenclature, classification, and mechanisms. Int J Legal Med.  2014;128(1):95-103. Epub 2013 Sep 24. Review.
  29. Scherer K, Harr T, Bach S, Bircher AJ. The role of iodine in hypersensitivity reactions to radio contrast media. Clin Exp Allergy. 2010;40(3):468-475.
  30. Gharekhanloo F, Torabian S. Comparison of allergic adverse effects and contrast enhancement between iodixanol and iopromide. Iran J Radiol. 2012;9(2):63-66. Epub 2012 Jun 30.
  31. Hawkins IF Jr, Mladinich CR, Storm B, et al. Short-term effects of selective renal arterial carbon dioxide administration on the dog kidney. J Vasc Interv Radiol. 1994;5(1):149-154.
  32. Davenport MS, Khalatbari S, Cohan RH, Ellis JH. Contrast medium-induced nephrotoxicity risk assessment in adult inpatients: a comparison of serum creatinine level and estimated glomerular filtration rate-based screening methods. Radiology. 2013;269(1):92-100. Epub 2013 Apr 11.
  33. McDonald JS, McDonald RJ, Comin J, et al. Frequency of acute kidney injury following intravenous contrast medium administration: a systematic review and meta-analysis. Radiology. 2013;267(1):119-128. Epub 2013 Jan 14.
  34. Martin-Paredero V, Dixon SM, Baker JD, et al. Risk of renal failure after major angiography. Arch Surg. 1983;118(12):1417-1420.
  35. Briguori C, Quintavalle C, De Micco F, Condorelli G. Nephrotoxicity of contrast media and protective effects of acetylcysteine. Arch Toxicol. 2011;85(3):165-173.
  36. Awal A, Ahsan SA, Siddique MA, et al. Effect of hydration with or without n-acetylcysteine on contrast induced nephropathy in patients undergoing coronary angiography and percutaneous coronary intervention. Mymensingh Med J. 2011;20(2):264-269.
  37. Oliva VL, Denbow N, Thérasse E, et al. Digital subtraction angiography of the abdominal aorta and lower extremities: carbon dioxide versus iodinated contrast material. J Vasc Interv Radiol. 1999;10(6):723-731.
  38. Weaver FA, Pentecost MJ, Yellin AE, Davis S, Finck E, Teitelbaum G. Clinical applications of carbon dioxide/digital subtraction arteriography. J Vasc Surg. 1991;13(2):266-272, discussion 272-273.
  39. Rolland Y, Duvauferrier R, Lucas A, et al. Lower limb angiography: a prospective study comparing carbon dioxide with iodinated contrast material in 30 patients. Am J Roentgenol. 1998;171(2):333-337.
  40. Seeger JM, Self S, Harward TR, Flynn TC, Hawkins IF. Carbon dioxide gas as an arterial contrast agent. Ann Surg. 1993;217(6):688-698; discussion 697-698.
  41. Weaver FA, Pentecost MJ, Yellin AE. Carbon dioxide digital subtraction arteriography: a pilot study. Ann Vasc Surg. 1990;4(5):437-441.
  42. Miller FJ, Mineau DE, Koehler PR, et al. Clinical intra-arterial digital subtraction imaging (use of small volumes of iodinated contrast material or carbon dioxide). Radiology. 1983;148(1):273-278.
  43. Hawkins IF, Caridi JG. Carbon dioxide digital subtraction angiography: 26-year experience at the University of Florida. Eur Radiol. 1998;8(3):391-402.

___________________________________________________________

From the 1Interventional Cardiology Department, Policlinico di Monza, Monza, Italy; 2Biostatistics Unit, Policlinico di Monza – Monza, Italy; and 3Physic Department, Università degli Studi di Bologna, Bologna, Italy.

Disclosure: The authors have completed and returned the ICMJE Form for Disclosure of Potential Conflicts of Interest. The authors report no conflicts of interest regarding the content herein.

Manuscript submitted April 10, 2014, provisional acceptance given May 15, 2014, final version accepted July 17, 2014.

Address for correspondence: Filippo Scalise, MD, FACC, FESC, FSCAI, Interventional Cardiology Department, Policlinico di Monza, Via Amati 111, 20900 Monza, Italy. Email: filippo.scalise@policlinicodimonza.it


Advertisement

Advertisement

Advertisement