AngioJet Rheolytic Thrombectomy During Rescue PCI for Failed
Thrombolysis: A Single-Center Experience
Methods
Study population. Two hundred and fourteen consecutive patients were transferred to Good Samaritan Hospital in Los Angeles, California from January 2000 through October 2004 to undergo rescue PCI for failed thrombolysis. The definition of failed thrombolysis was established on clinical grounds and made by the referring physician. All patients had ongoing chest pain, continued ST-segment elevation, hemodynamic instability or malignant ventricular arrhythmias and required transfer for emergent coronary angiography. All patients (n = 214) underwent emergent coronary angiography within 6 hours of arrival. Thirty-two patients (15%) underwent AngioJet RT because of intracoronary thrombus per physician discretion (RT group, n = 32). A group of 32 control patients were identified by matching patients based on infarct related artery (IRA) location and initial thrombolysis in MI (TIMI) flow grade (control group, n = 32).
Interventional procedures and definitions. Cardiogenic shock was defined as systolic blood pressure < 80 mmHg despite fluid boluses or dependence on vasopressor agents. All patients were treated with thrombolytic therapy prior to transfer (streptokinase 2%; tenecteplase 65%; tissue-type plasminogen activator 26%; retevase 7%). At the time of PCI, heparin was administered following arterial access to maintain an activated clotting time (ACT) of 200 to 250 if a glycoprotein (GP) IIb/IIIa inhibitor was used and 250 to 300 if a GP IIb/IIIa inhibitor was not used. Coronary angioplasty and stent placement were performed using standard techniques. AngioJet rheolytic thrombectomy was performed if significant thrombus was present at the discretion of the operator. Vascular access sheaths were removed 10 to 12 hours following the procedure if the fibrinogen level was greater than 200. After PCI, all patients received 325 mg aspirin and either ticlopidine 250 mg twice daily or clopidogrel 75 mg once per day for 1 month if a bare metal stent was placed and for 6 months if a drug-eluting stent was placed.
Recurrent ischemia was defined as anginal chest pain and electrocardiogram (ECG) changes (ST-segment depression or T-wave changes) occurring after the index PCI procedure and not accompanied by elevated cardiac enzymes. Recurrent MI was defined as anginal chest pain and ST-segment re-elevation on ECG occurring after the index PCI procedure associated with elevated cardiac enzymes. No reflow was defined as transient or permanent reduction of TIMI flow grade to 0 or 1 from a higher flow grade, at any point during the procedure. Bleeding events were classified as severe or moderate according to the definitions of the Global Use of Strategies to Open Occluded Coronary Arteries (GUSTO 2b) Trial.5 Severe bleeding was defined as intracranial bleeding or bleeding resulting in hemodynamic compromise. Moderate bleeding was defined as bleeding that required blood transfusion but that did not lead to hemodynamic compromise. Major adverse cardiac events (MACE) were defined as the composite of death, recurrent MI or ischemia, emergent coronary artery bypass grafting (CABG) or repeat target vessel revascularization (TVR). Emergent CABG was defined as CABG within 24 hours of an unsuccessful PCI. Angiographic success was defined as restoration of TIMI 3 flow in the IRA with £ 50% residual stenosis. Clinical success was defined as angiographic success with freedom from MACE.
Angiographic analysis. Angiograms were reviewed by two investigators blinded to clinical information and outcome. The initial injection of the IRA was used to determine initial TIMI flow grade and final TIMI flow grade was assessed following guidewire removal.6 Thrombus grade was evaluated by the methods of the TIMI study group.7 TIMI frame count (TFC) was measured at the initial and final angiogram by the methods of Gibson et al.8 The culprit coronary lesion was classified by American College of Cardiology (ACC)/American Heart Association (AHA) criteria.9 Statistical analysis. Analyses were performed using SAS® 3.0 Statistical Software (SAS Institute Inc, Cary, North Carolina). Continuous variables are presented as means ± standard deviations and were compared using student’s t-test and frequencies were compared using the chi-square test. A p-value of < 0.05 was considered significant.
Results
Baseline characteristics. Characteristics of the study groups are shown in Table 1. In general, the groups were similar, except the RT group had a higher prevalence of males and the Control group had a higher prevalence of patients intubated prior to transfer. Time from thrombolytic therapy to angiography and the use of GP IIb/IIIa inhibitors, percutaneous transluminal coronary angioplasty and coronary stents also were similar between the groups. Length of implanted stent was similar among the two groups.
Angiographic findings. The prevalence of ACC/AHA type C lesion and initial percent stenosis were similar among the two groups (Table 2). Final angiography showed a trend for a higher prevalence of TIMI grade 3 flow in the RT compared to the Control group (94% vs. 78%, p = 0.07, respectively) (Figure 1). At initial angiography, there was a higher prevalence of combined TIMI thrombus grade 4 and 5 in the RT compared to the Control group (84% vs. 59%, p = 0.03, respectively), indicating a higher thrombus burden in the RT compared to the Control group prior to PCI. At final angiography, more patients in the RT group achieved TIMI thrombus grade 0 compared to those in the Control group (94% vs. 69%, p = 0.01, respectively) (Figure 1). Distal embolization and coronary perforation occurred equally in both groups (Table 2). No reflow was significantly reduced in the RT compared to the Control group (13% vs. 56%, p = 0.0002, respectively). TIMI frame count was similar between both groups at initial and at final angiography (Table 2). The improvement in TIMI frame count from initial to final angiography was similar between both groups: 48 ± 33 for the RT group and 46 ± 32 for the Control group (Figure 1).
Clinical outcome. Major bleeding events, stroke, need for emergent bypass surgery, recurrent MI, emergent TVR and death occurred equally in both groups (Table 3). More patients in the Control group developed renal failure compared to the RT group (19% vs. 3%, p = 0.05, respectively). The contrast volume and fluoroscopy times were similar among both groups (Table 2). Angiographic success showed a trend to be higher in the RT compared to the Control group (94% vs. 78%, p = 0.07, respectively). Clinical success was higher in the RT compared to the Control group (91% vs. 72%, p = 0.05, respectively).
Discussion
The treatment of thrombus-containing lesions during PCI remains problematic.10 A variety of devices have been reported to remove angiographically apparent coronary thrombus. The Pronto catheter (Vascular Solutions Inc, Minneapolis, Minnesota),11 allows manual aspiration of clot, but lacks a safety and/or efficacy trial of significant size in the coronary circulation. The X-sizer catheter system (EndiCOR Medical Inc., San Clemente, California)12 employs a mechanical clot removal mechanism and has been studied in the coronary circulation, but lacks FDA approval for coronary use. The AngioJet RT system performs a mechanical thrombectomy and has been validated as an effective method to remove thrombus in the setting of acute MI.13,14 In the current study, these observations were extended by characterizing the use of AngioJet RT in a consecutive series of patients undergoing rescue PCI for failed thrombolysis. Patients undergoing AngioJet RT achieved more effective thrombus removal as indicated by a higher prevalence of TIMI 0 thrombus grade at final angiography. This thrombus removal superiority occurred despite the unfair bias of more extensive thrombus burden in the AngioJet RT group at initial angiography as demonstrated by a higher prevalence of TIMI thrombus grades 4 and 5, compared to the Control group. GP IIb/IIIa inhibitors have also been advocated as a treatment for thrombus containing lesions.7,15 However, GP IIb/IIIa inhibitors have not been specifically studied in the setting of rescue PCI for failed thrombolysis16 and their use may be associated with increased bleeding complications.17
The use of AngioJet RT was associated with a significantly lower incidence of the no reflow phenomenon, higher clinical success and a trend toward higher angiographic success. Reducing the occurrence of no reflow is clearly beneficial during rescue PCI for failed thrombolysis, given the critical nature of these patients and the detrimental effects of no reflow on clinical outcome.18–20 In addition, there were no cases of coronary perforation and major bleeding events were similar compared to the control group, indicating the safety of AngioJet RT during rescue PCI.
Two recently reported randomized trials have sought to validate rescue PCI as a viable clinical strategy following failed thrombolysis.21,22 Both the Middlesbrough Early Revascularization to Limit Infarction (MERLIN) Trial and the Rescue Angiography after Failed Thrombolysis (REACT) Trial were randomized studies which found a benefit for rescue PCI compared to conservative therapy. The findings of the REACT Trial were more compelling as both recurrent MI and the combined endpoint for rescue PCI were reduced. In contrast, the MERLIN Trial only found a benefit of rescue PCI for the combined endpoint that was driven primarily by a reduction in revascularization rates. Differences in trial design may partially explain these findings as the use of streptokinase as the thrombolytic agent was significantly lower in REACT compared to the MERLIN Trial (59% vs. 96%). In addition, there was a significantly higher use of coronary stents and GP IIb/IIIa agents in the REACT Trial. Our study attempts to gain additional insight into whether further technical refinement during rescue PCI, with mechanical thrombus removal using AngioJet RT, might yield further improvements in clinical outcome. It was found that AngioJet RT achieved more effective thrombus removal with the achievement of TIMI thrombus grade 0 in 94% of patients (compared to 69% in the Control group) and markedly reduced the incidence of no-reflow to 13% (compared to 59% in the Control group). While the results using AngioJet RT during rescue PCI are encouraging, the current study is limited by a relatively small subgroup and larger trials will be required to verify these results.
Study Limitations
This report represents a single-center, observational, prospective study of consecutive patients undergoing rescue PCI for failed thrombolysis treated at 1 institution by a single group of interventional cardiologists. The use of AngioJet RT was not pre-specified, but was done at operator discretion for thrombus-containing lesions.
Conclusions
AngioJet RT with rescue PCI for failed thrombolysis in this study was safe and effectively removed thrombus, reduced the occurrence of no re-flow and tended to improve procedural outcomes. Physicians should consider AngioJet RT in the setting of angiographic thrombus during rescue PCI.
References
- Ellis SG, Roubin GS, King SB, III, et al. Angiographic and clinical predictors of acute closure after native vessel coronary angioplasty. Circulation 1988;77:372–379.
- Singh M, Berger PB, Ting HH, et al. Influence of coronary thrombus on outcome of percutaneous coronary angioplasty in the current era (the Mayo Clinic experience). Am J Cardiol 2001;88:1091–1096.
- Silva JA, Ramee SR, Cohen DJ, et al. Rheolytic thrombectomy during percutaneous revascularization for acute myocardial infarction: Experience with the AngioJet catheter. Am Heart J 2001;141:353–359.
- Kuntz RE, Baim DS, Cohen DJ, et al. A trial comparing rheolytic thrombectomy with intracoronary urokinase for coronary and vein graft thrombus (the Vein Graft AngioJet Study [VeGAS 2]). Am J Cardiol 2002;89:326–330.
- An international randomized trial comparing four thrombolytic strategies for acute myocardial infarction. The GUSTO investigators. N Engl J Med 1993;329:673–682.
- Chesebro JH, Knatterud G, Roberts R, et al. Thrombolysis in Myocardial Infarction (TIMI) Trial, Phase I: A comparison between intravenous tissue plasminogen activator and intravenous streptokinase. Clinical findings through hospital discharge. Circulation 1987;76:142–154.
- Gibson CM, de Lemos JA, Murphy SA, et al. Combination therapy with abciximab reduces angiographically evident thrombus in acute myocardial infarction: A TIMI 14 substudy. Circulation 2001;103:2550–2554.
- Gibson CM, Cannon CP, Daley WL, et al. TIMI frame count: A quantitative method of assessing coronary artery flow. Circulation 1996;93:879–888.
- Ryan TJ, Faxon DP, Gunnar RM, et al. Guidelines for percutaneous transluminal coronary angioplasty. A report of the American College of Cardiology/American Heart Association Task Force on Assessment of Diagnostic and Therapeutic Cardiovascular Procedures (Subcommittee on Percutaneous Transluminal Coronary Angioplasty). Circulation 1988;78:486–502.
- Topaz O. On the hostile massive thrombus and the means to eradicate it. Catheter Cardiovasc Interv 2005;65:280–281.
- Lim MJ, Reis L, Ziaee A, Kern MJ. Use of a new thrombus extraction catheter (the Pronto) in the treatment of acute myocardial infarction. J Interv Cardiol 2005;18:189–192.
- Lefevre T, Garcia E, Reimers B, et al. X-sizer for thrombectomy in acute myocardial infarction improves ST-segment resolution: Results of the X-sizer in AMI for negligible embolization and optimal ST resolution (X AMINE ST) trial. J Am Coll Cardiol 2005;46:246–252.
- Silva JA, Ramee SR, Cohen DJ, et al. Rheolytic thrombectomy during percutaneous revascularization for acute myocardial infarction: Experience with the AngioJet catheter. Am Heart J 2001;141:353–359.
- Nakagawa Y, Matsuo S, Yokoi H, et al. Stenting after thrombectomy with the AngioJet catheter for acute myocardial infarction. Cathet Cardiovasc Diagn 1998;43:327–330.
- Muhlestein JB, Karagounis LA, Treehan S, Anderson JL. “Rescue” utilization of abciximab for the dissolution of coronary thrombus developing as a complication of coronary angioplasty. J Am Coll Cardiol 1997;30:1729–1734.
- Petronio AS, De CM, Rossini R, et al. Role of platelet glycoprotein IIb/IIIa inhibitors in rescue percutaneous coronary interventions. Ital Heart J 2004;5:114–119.
- Jong P, Cohen EA, Batchelor W, et al. Bleeding risks with abciximab after full-dose thrombolysis in rescue or urgent angioplasty for acute myocardial infarction. Am Heart J 2001;141:218–225.
- Morishima I, Sone T, Okumura K, et al. Angiographic no-reflow phenomenon as a predictor of adverse long-term outcome in patients treated with percutaneous transluminal coronary angioplasty for first acute myocardial infarction. J Am Coll Cardiol 2000;36:1202–1209.
- Feld H, Lichstein E, Schachter J, Shani J. Early and late angiographic findings of the “no-reflow” phenomenon following direct angioplasty as primary treatment for acute myocardial infarction. Am Heart J 1992;123:782–784.
- Romano M, Buffoli F, Lettieri C, et al. No reflow in patients undergoing primary angioplasty for acute myocardial infarction at high risk: Incidence and predictive factors. Minerva Cardioangiol 2005;53:7–14.
- Sutton AG, Campbell PG, Graham R, et al. A randomized trial of rescue angioplasty versus a conservative approach for failed fibrinolysis in ST-segment elevation myocardial infarction: The Middlesbrough Early Revascularization to Limit INfarction (MERLIN) trial. J Am Coll Cardiol 2004;44:287–296.
- Gershlick AH, Stephens-Lloyd A, Hughes S, et al. Rescue angioplasty after failed thrombolytic therapy for acute myocardial infarction. N Engl J Med 2005;353:2758–2768.