Poster LR-042

Using Standardized Testing to Delineate How Materials and Construction Impact Performance Characteristics of Support Surfaces to Enhance Objective Clinical Decision-making for Pressure Injury Prevention

Kristen Thurman, PT, MPT, CWS

Symposium on Advanced Wound Care Spring Spring 2022

Introduction: Clinicians are challenged with having objective data to choose support surfaces that will offer the most ideal clinical performance for patient populations. Support surfaces are designed with different materials and constructed in various ways to provide therapeutic value. Using laboratory Support Surface Standards testing could assist with understanding the performance differences between surfaces providing more objective data for clinicians to match performance characteristics to patient need.

Methods: Support surfaces with various cover materials and construction were selected to test using the 2019 American National Standard for Support Surfaces Standardized tests for Immersion.1 A foam mattress with a nylon cover, a foam-filled air cell surface with a polycarbonate cover, and a 10” air surface with a polycarbonate cover were selected. The Immersion and Envelopment Standardized Tests, Section 2 and 6, were performed on each surface.

Results: The immersion results were 26.4% for the foam/nylon cover, 30% for the foam-filled cells/polycarbonate cover, and 61% for the 10” air cells/polycarbonate cover. The foam/nylon surface had the least amount of immersion, whereas the air/polycarbonate surface had the most immersion.

Discussion: The materials and construction of the support surfaces did influence the performance characteristic of immersion. Clinicians looking for surfaces with higher immersion, for example with immobile patient populations where more pressure redistribution is desired, may consider surfaces constructed with air cells and a polycarbonate cover over foam surfaces with a nylon cover. Patients who are working in rehabilitation may benefit from less immersion, therefore one of the foam surfaces may be a better choice. Every support surface, with a varying design and cover material, will have a different outcome, as evidenced by Call and Capunay. They reported that surfaces had varying results of immersion from 47.3% (a powered air surface) to 31.4% (a self-adjusting technology air surface).2 Obtaining these standardized test results will give clinicians objective data in which to compare performance characteristics of support surfaces.

Conclusion: Differences in support surface design and materials influence performance characteristics that can affect patient outcomes. Standardized testing allows for objective comparisons of support surface performance characteristics for clinicians to make evidenced-based choices for specific patient populations.

References

1. American National Standards Institute/Rehabilitation Engineering and Assistive Technology Society of North America. SS-1-2019 RESNA Standard for Support Surfaces—Volume 1: Requirements and Test Methods for Full Body Support Surfaces. Arlington, VA: Rehabilitation Engineering and Assistive Technology Society of North America; 2019.2. www.stryker.com/content/dam/stryker/campaigns/acute-care/resources/IsoTour%20White%20Paper%20Summary%20Poster%20(1).pdf Call E, Capunay C. Standardized testing to evaluate the microclimate, immersion, and envelopment capabilities of a support surface. Accessed December 17, 2021.